Dual Test Gas Pulmonary Diffusing Capacity Measurement During Exercise in Humans Using the Single-Breath Method
Published on: February 2, 2024
Long-term intersession variability for single-breath diffusing capacity
Matthew J Hegewald1, Robert L Jensen, John G Teeter
1Division of Pulmonary and Critical Care Medicine, University of Utah School of Medicine and Intermountain Medical Center, Salt Lake City/Murray, Utah, USA.
Long-term diffusing capacity (DL(CO)) variability in biocontrols was characterized. Deviations greater than 12% or 3.0 units from the average of the first six tests suggest equipment quality control issues.
Area of Science:
- Pulmonary Function Testing
- Biomedical Engineering
- Quality Control in Healthcare
Background:
- Assessing long-term diffusing capacity (DL(CO)) variability is crucial for equipment quality control and patient management.
- Previous research has not reported on long-term DL(CO) variability.
Purpose of the Study:
- To characterize the long-term variability of DL(CO) in a biocontrol cohort.
- To compare different methods for selecting a target DL(CO) value for variability assessment.
Main Methods:
- Longitudinal DL(CO) monitoring of 288 biocontrols over up to 5 years, with twice-monthly testing.
- Standardized DL(CO) measurement technique used throughout the study.
- Variability analysis using three distinct target value selection methods.
Main Results:
- The 90th percentile for mean intersession DL(CO) change ranged from 10.9% to 15.8% (2.6-4.1 units).
- Lowest variability was observed when using the mean of all tests as the target; highest variability occurred when using the baseline DL(CO).
- The average of the first six DL(CO) tests accurately estimated the mean DL(CO) value, yielding a 90th percentile change of 12.3% (3.0 units).
Conclusions:
- DL(CO) deviations exceeding 12% or 3.0 units from the average of the first six tests indicate potential instrument quality control issues.
- Instruments showing such deviations require careful evaluation before further patient testing.
- Variability remained stable over time and was not significantly associated with demographic factors.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
08:23Quantitative Measure of Lung Structure and Function Obtained from Hyperpolarized Xenon Spectroscopy
Published on: November 10, 2023
Related Concept Videos
External and Internal Respiration
Respiratory Volumes and Capacities I
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Respiratory Volumes and Capacities