Validity and reliability of central blood pressure estimated by upper arm oscillometric cuff pressure
Rachel E D Climie1, Martin G Schultz, Sonja B Nikolic
1Menzies Research Institute Tasmania, University of Tasmania, Hobart, Australia.
Background:
Noninvasive central blood pressure (BP) independently predicts mortality, but current methods are operator-dependent, requiring skill to obtain quality recordings. The aims of this study were first, to determine the validity of an automatic, upper arm oscillometric cuff method for estimating central BP (O(CBP)) by comparison with the noninvasive reference standard of radial tonometry (T(CBP)). Second, we determined the intratest and intertest reliability of O(CBP).
Methods:
To assess validity, central BP was estimated by O(CBP) (Pulsecor R6.5B monitor) and compared with T(CBP) (SphygmoCor) in 47 participants free from cardiovascular disease (aged 57 ± 9 years) in supine, seated, and standing positions. Brachial mean arterial pressure (MAP) and diastolic BP (DBP) from the O(CBP) device were used to calibrate in both devices. Duplicate measures were recorded in each position on the same day to assess intratest reliability, and participants returned within 10 ± 7 days for repeat measurements to assess intertest reliability.
Results:
There was a strong intraclass correlation (ICC = 0.987, P < 0.001) and small mean difference (1.2 ± 2.2 mm Hg) for central systolic BP (SBP) determined by O(CBP) compared with T(CBP). Ninety-six percent of all comparisons (n = 495 acceptable recordings) were within 5 mm Hg. With respect to reliability, there were strong correlations but higher limits of agreement for the intratest (ICC = 0.975, P < 0.001, mean difference 0.6 ± 4.5 mm Hg) and intertest (ICC = 0.895, P < 0.001, mean difference 4.3 ± 8.0 mm Hg) comparisons.
Conclusions:
Estimation of central SBP using cuff oscillometry is comparable to radial tonometry and has good reproducibility. As a noninvasive, relatively operator-independent method, O(CBP) may be as useful as T(CBP) for estimating central BP in clinical practice.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: