Measurement uncertainty in single, double and triple isotope dilution mass spectrometry
1BAM Federal Institute for Materials Research and Testing, Unter den Eichen 87, 12205, Berlin, Germany. jochen.vogl@bam.de
Abstract:
Triple IDMS has been applied for the first time to the quantification of element concentrations. It has been compared with single and double IDMS obtained on the same sample set in order to evaluate the advantages and disadvantages of triple IDMS over single and double IDMS as an analytical reference procedure. The measurement results of single, double and triple IDMS are indistinguishable, considering rounding due to the individual measurement uncertainties. As expected, the relative expanded uncertainties (k = 2) achieved with double IDMS (0.08%) are dramatically smaller than those obtained with single IDMS (1.4%). Triple IDMS yields the smallest relative expanded uncertainties (k = 2, 0.077%) unfortunately at the expense of a much higher workload. Nevertheless triple IDMS has the huge advantage that the isotope ratio of the spike does not need to be determined. Elements with high memory effects, highly enriched spikes or highest metrological requirements may be typical applications for triple IDMS.