Characteristics of x-ray beams in two commercial multidetector computed tomography simulators: Monte Carlo
Sangroh Kim1, Haijun Song, Benjamin Movsas
1Department of Radiation Oncology, Henry Ford Health System, Detroit, Michigan 48202, USA. skim4@hfhs.org
Purpose:
As multidetector computed tomography (MDCT) scanning is routinely performed for treatment planning in radiation oncology, understanding the characteristics of the MDCT x-ray beam is essential to accurately estimate patient dose. The purpose of this study is to characterize the x-ray beams of two commercial MDCT simulators widely used in radiation oncology by Monte Carlo (MC) simulations.
Methods:
X-ray tube systems of two wide bore MDCT scanners (GE LightSpeed RT 4 and Philips Brilliance Big Bore) were modeled in the BEAMNRC/EGSNRC MC system. All the tube components were modeled from targets to bowtie filters. To validate our MC models, the authors measured half-value layers (HVL) using aluminum sheets and multifunctional radiation detectors and compared them to those obtained from MC simulations for 120 kVp beams. The authors also compared x-ray spectra obtained from MC simulation to the data provided by manufacturers. Additionally, lateral/axial beam profiles were measured in-air using radiochromic films and compared to the MC results. To understand the scatter effect, the authors also derived the scatter-to-primary energy fluence ratio (SPR) profiles and calculated the total SPR for each CT system with the CT dose index (CTDI) head and body phantoms using the BEAMNRC system.
Results:
The authors found that the HVL, x-ray spectrum and beam profiles of the MC simulations agreed well with the manufacturer-specified data within 1%-10% on average for both scanners. The total SPR were ranged from 7.8 to 13.7% for the head phantom and from 10.7 to 18.9% for the body phantom.
Conclusions:
The authors demonstrate the full MC simulations of two commercial MDCT simulators to characterize their x-ray beams. This study may be useful to establish a patient-specific dosimetry for the MDCT systems.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies III: Computed Tomography