Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
Published on: April 29, 2011
The flow-time waveform predicts respiratory system resistance and compliance
Boulos S Nassar1, Nicole D Collett, Gregory A Schmidt
1Division of Pulmonary, Critical Care, and Occupational Medicine, Department of Internal Medicine, University of Iowa Carver College of Medicine, Iowa City, IA 52242, USA. boulos-nassar@uiowa.edu
Lung compliance and resistance can be estimated during pressure assist-control ventilation (PACV), correlating well with values from volume assist-control ventilation (VACV). This finding aids clinical management of mechanically ventilated patients.
Area of Science:
- Respiratory Mechanics
- Mechanical Ventilation
Background:
- Accurate assessment of lung compliance and resistance is crucial for effective clinical management of mechanically ventilated patients.
- These parameters are routinely measured during volume assist-control ventilation (VACV).
Purpose of the Study:
- To investigate if lung compliance and resistance can be reliably estimated during pressure assist-control ventilation (PACV).
- To compare these estimations with values obtained during VACV.
Main Methods:
- Collected data from 12 mechanically ventilated human subjects.
- Measured respiratory parameters during both VACV and PACV modes.
- Calculated inspiratory resistance and compliance during PACV by extrapolating the linear component of the pressure-time waveform.
Main Results:
- Mean compliance during PACV (37.06 ± 15.65 mL/cm H2O) closely matched VACV values (36.93 ± 12.18 mL/cm H2O).
- Mean inspiratory resistance during PACV (15.17 ± 5.14 cm H2O/L/s) showed strong correlation with VACV values (12.50 ± 2.99 cm H2O/L/s), with R² of 0.73 and 0.51 respectively.
- A strong correlation was observed between compliance and inspiratory resistance calculated during PACV versus VACV.
Conclusions:
- The inspiratory flow waveform during PACV is linear and contains information about respiratory system resistance and compliance.
- Estimated lung compliance and resistance during PACV correlate well with conventionally measured values during VACV.
- These findings support the use of PACV for estimating key respiratory mechanics parameters.
Related Concept Videos
Application of Integration: Problem Solving
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...