Oscillometric sphygmomanometers: a critical appraisal of current technology
1Department of Medical Physics, Crosshouse Hospital, Kilmarnock, Scotland, UK. john.amoore@aapct.scot.nhs.uk
Aim:
Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy.
Methods:
A block diagram is presented covering the components from cuff to algorithm. The oscillometric waveform is described, considering factors that can alter its shape. Methods used to assess accuracy, including the potential use of simulators, are described.
Results And Discussion:
The block diagram focuses attention on cuff, amplifier, signal processing and algorithm. The importance of correct cuff size is emphasized. Accuracy can be affected by the extraction of the oscillometric pulses and the interpolation to compensate for higher deflation rates. Modern electronic amplifiers are assumed to be stable and do not drift, an assumption largely untested. Crucial to accuracy is the algorithm, but there is no standard algorithm and limited theoretical basis, leading to significant measurement errors in groups of patients, even by approved devices. The causes are not well understood, but differences in oscillometric waveform shape between patient groups have been observed and may explain auscultatory-oscillometric differences. The ability of theoretical models to explain the effects of arterial stiffness on BP measurements is discussed. Validation remains statistical though steps have been taken to improve it.
Conclusion:
The indirect nature of BP measurement poses particularly problems for ensuring accuracy. Critical assessment has done much to improve standards, but a solid theoretical understanding of the technique has not been formulated and further work is required.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: