Efficacy of lower-body shielding in computed tomography fluoroscopy-guided interventions
Andreas H Mahnken1, Martin Sedlmair, Christine Ritter
1Department of Diagnostic and Interventional Radiology, RWTH Aachen, 52074, Aachen, Germany. mahnken@rad.rwth-aachen.de
Purpose:
Computed tomography (CT) fluoroscopy-guided interventions pose relevant radiation exposure to the interventionalist. The goal of this study was to analyze the efficacy of lower-body shielding as a simple structural method for decreasing radiation dose to the interventionalist without limiting access to the patient.
Material And Methods:
All examinations were performed with a 128-slice dual source CT scanner (12 × 1.2-mm collimation; 120 kV; and 20, 40, 60, and 80 mAs) and an Alderson-Rando phantom. Scatter radiation was measured with an ionization chamber and a digital dosimeter at standardized positions and heights with and without a lower-body lead shield (0.5-mm lead equivalent; Kenex, Harlow, UK). Dose decreases were computed for the different points of measurement.
Results:
On average, lower-body shielding decreased scatter radiation by 38.2% within a 150-cm radius around the shielding. This decrease is most significant close to the gantry opening and at low heights of 50 and 100 cm above the floor with a maximum decrease of scatter radiation of 95.9% close to the scanner's isocentre. With increasing distance to the gantry opening, the effect decreased. There is almost no dose decrease effect at ≥150 above the floor. Scatter radiation and its decrease were linearly correlated with the tube current-time product (r (2) = 0.99), whereas percent scatter radiation decrease was independent of the tube current-time product.
Conclusion:
Lower-body shielding is an effective way to decrease radiation exposure to the interventionalist and should routinely be used in CT fluoroscopy-guided interventions.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography