Related Experiment Video
Updated: May 25, 2026

Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Ultrahigh sensitivity assays for human cardiac troponin I using TiO2 nanotube arrays
Piyush Kar1, Archana Pandey, John J Greer
1Department of Electrical and Computer Engineering, University of Alberta, Edmonton, AB, Canada.
Abstract:
Rapid, highly sensitive troponin assays for the analysis of serum at the point-of-care are particularly desirable for the effective treatment of myocardial infarction (MYI). TiO(2) nanotube arrays constitute a low cost, high surface area, semiconducting architecture with great promise for biosensing applications due to their compatibility with multiple detection techniques. Using TiO(2) nanotube arrays functionalized with highly robust and ordered carboxyalkylphosphonic acid self-assembled monolayers, we have developed a simple and highly sensitive fluorescence immunoassay which can detect concentrations of human cardiac troponin I as low as 0.1 pg ml(-1) without the use of enzymatic amplification. Varying the morphological parameters of the nanotube arrays allows tuning the detection range over 6 orders of magnitude of the troponin concentration from 0.1 pg ml(-1)-100 ng ml(-1).
More Related Videos
08:58Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay