Imaging of the Microstructural Failure Mechanism in the Human Hip
Published on: September 29, 2023
Cementation and interface analysis of early failure cases after hip-resurfacing arthroplasty
Matthias Krause1, Stefan Breer, Michael Hahn
1Department of Osteology and Biomechanics, University Medical Center Hamburg-Eppendorf, Lottestraße 59, 22529, Hamburg, Germany.
Purpose:
The use of inappropriate cementation techniques has been suggested as an adverse factor for the long-term survival of hip-resurfacing arthroplasty. Inadequate initial fixation, thermal osteonecrosis and interface biological reactions are possible causes of failure. We analysed morphological changes associated with the cementation technique in a large collection of retrieved femoral components.
Methods:
One hundred and fifty femoral components (mean time to failure of 8.3 months±11.0) obtained at revision surgery were analysed morphometrically and histopathologically. Cement mantle and penetration were quantified in six different regions of interest. Histopathological analysis of the bone-cement interface was performed on undecalcified processed bone tissue.
Results:
The vast majority of the cases differed substantially from laboratory-based cement-penetration depth recommendations. Fifty-nine cases had a fibrous membrane at the cement-bone interface. This membrane was significantly thicker in cases with osteonecrosis compared to cases viable bone.
Conclusions:
Our results demonstrate that most failures were cemented inappropriately. We suggest that poor cementation was an important adverse factor; however, the cause of the failures was obviously multifactorial. The thickness of the fibrous membrane at the cement-bone interface differed significantly between cases with osteonecrosis and specimens with viable bone tissue.