Humeral head resurfacing in central bone defects: in vitro stability of different implants with increasing defect
Philip Kasten1, Christine Neubrech, Patric Raiss
1Division of Shoulder/Elbow Surgery and Sports Medicine, Department of Orthopaedic Surgery, University Hospital Carl Gustav Carus Dresden, Fetscherstrasse 74, 01307 Dresden, Germany.
Abstract:
We examined the rotary motions of two distinct cementless surface replacement arthroplasties of the shoulder that were implanted on humeri with central spherical bone defects of 8%, 17%, and 37% of the head volume (n = 5 each). Rotary motions were measured under a cyclic torque application and translated into relative micromotions. Implant A with a perforated central crown had micromotions < 150 µm in all bony defects and during all simulated shoulder activities. Implant B with a central tapered tri-fin pin had no micromotions > 150 µm in defects of 0% and 8% during a strenuous activity like lifting 10 kg, but did exhibit micromotion > 150 µm in 40% of the experiments in defects of 17% and 37%, which could impair bony ingrowth. Implant B displayed a significant increase in micromotions for defects of 8% and 17% (p < 0.05). Our results suggest that implant A could be used without risk in spherical head defects up to 37% of volume, even in strenuous shoulder activities immediately after its implantation. Implant B, however, is recommended in spherical defects starting from 8% and only during light shoulder activities until bony integration of the implant has occurred.