Related Experiment Video
Updated: May 25, 2026

Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
A novel device for endovascular native aortic valve resection for transapical transcatheter aortic valve implantation
Parla Astarci1, David Glineur, Gébrine Elkhoury
1Department of Cardiovascular and Thoracic Surgery, University Hospital Saint-Luc, Brussels, Belgium. parla.astarci@uclouvain.be
Abstract:
We developed a novel resection device to use during transapical transcatheter aortic valve implantation (TAVI) using a circular blade. We assessed the device in 15 human cadavers by transapical approach. After the resection, the aortic annulus was measured using standard probes. A careful examination of the aortic wall, left ventricular outflow tract, coronary ostia and mitral valve was performed using an endpoint checklist, developed specifically for the new device. The resection was successfully completed in 14 out of 15 (93%) cases. All the resected leaflets and debris have been successfully evaluated in 15 out of 15 (100%) cases. One case of a bicuspid valve had a prominent calcification of the median raphe. The resection tool could only perform a partial resection. The mean duration of the resection was 45 ± 30 s. The surrounding tissue examination did not reveal any injury to the anatomical structures. Endovascular resection of the native valve using transapical approach is feasible and effective. Further developments are necessary before the definitive clinical use during percutaneous aortic valve implantation.
