Related Experiment Video
Updated: May 25, 2026

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Feasibility evaluation of a motion detection system with face images for stereotactic radiosurgery
Takuya Yamakawa1, Koichi Ogawa, Hitoshi Iyatomi
1Department of Applied Informatics, Faculty of Science and Engineering, Hosei University, Kajinocho, Koganei, Tokyo 184-8584, Japan.
Abstract:
In stereotactic radiosurgery we can irradiate a targeted volume precisely with a narrow high-energy x-ray beam, and thus the motion of a targeted area may cause side effects to normal organs. This paper describes our motion detection system with three USB cameras. To reduce the effect of change in illuminance in a tracking area we used an infrared light and USB cameras that were sensitive to the infrared light. The motion detection of a patient was performed by tracking his/her ears and nose with three USB cameras, where pattern matching between a predefined template image for each view and acquired images was done by an exhaustive search method with a general-purpose computing on a graphics processing unit (GPGPU). The results of the experiments showed that the measurement accuracy of our system was less than 0.7 mm, amounting to less than half of that of our previous system.