Related Experiment Video
Updated: May 25, 2026

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
The effect of electrocardiographic lead choice on RR time series
M A García-González1, J Ramos-Castro, M Fernández-Chimeno
1Group of Biomedical and Electronic Instrumentation, Department of Electronic Engineering, Technical University of Catalonia, Barcelona 08034, Spain. miquel.angel.garcia@upc.edu
Abstract:
Results of heart rate variability analysis depend on the quality of the initial RR time series that is measured only in one lead of the ECG. This work shows that RR time series can subtly change from lead to lead so the choice of the analyzed lead is another source of uncertainty. The standard deviation of the differences of two RR time series obtained from different leads can change from 0.5 ms to more than 20 ms depending on the amount of noise, the morphological changes of the QRS complexes, the strategies of fiducial point determination and the measured subject. This source of uncertainty is in healthy subjects greater than that associated to the sampling frequency of the ECG for sampling frequencies greater than 400 Hz.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Dysrhythmias V: Evaluating Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...