The development of feedback monitoring device for CPR
Yeong-Tak Song1, Youngjoon Chee
1Department of Bioengineering, University of California, San Diego, La Jolla, CA, USA. tagy25@naver.com
Abstract:
CPR (Cardio-Pulmonary Resuscitation) is known as the most basic aid in emergency situations. For successful CPR, the chest compression depth, cycle, and compressing point are important factors. In I.C.U.'s (Intensive Care Unit) and E.R.'s (Emergency Room), monitoring devices are used to monitor the chest compressions correctly. These devices use accelerometers or pressure sensors. Because the mattress under the patient compresses together, these devices overestimate the compression depth. To overcome this problem, two accelerometers are used in this study, one is on the chest, and the other is between the back of the patient and the mattress. The measurement setup and algorithm to estimate the compression depth are explained. According to the experiment with the mannequin, when CPR is done on a mattress, the actual compression depth was 43.7 mm (S.D. 1.93 mm). The estimated compression depth was 61.4mm (S.D. 1.87 mm) when using an acceleration sensor. This includes the depth of compression of the mattress. When we use two acceleration sensors, estimated compression depth is 44.6mm (S.D. 1.59 mm), which is similar to the actual compression depth. In conclusion, the dual accelerometer gives more accurate estimated compression depth than conventional devices.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation I: Adult
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring
Cardiopulmonary Resuscitation II: ACLS Airway Management