A bio-impedance probe to assess liver steatosis during transplant surgery
Penny Probert Smith1, Fusheng You, Thomas Vogel
1Dept of Engineering Science, Institute of Biomedical Engineering, University of Oxford, UK. penny.smith@eng.ox.ac.uk
Abstract:
This work addresses the design of a bioimpedance probe to assess steatosis on the exposed liver in the donor during liver transplant surgery. Whereas typically bioimpedance uses needle probes to avoid surface effects, for clinical reasons a non-penetrative probe is required. In addition the need to ensure that the measurement is representative of the bulk tissue suggests a larger probe than is normally used to ensure a sufficiently large measurement volume. Using a simple model, simulations and tests on bovine liver, this paper investigates the relationship between probe dimensions and depth of measurement penetration and investigates the accuracy which might be expected in a configuration suitable for use in the operating theatre on intact but exposed livers. A probe using ECG electrodes is proposed and investigated.