Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Initial study on pulse wave velocity acquired from one hand using two synchronized wireless reflectance pulse
1Department of Electrical & Computer Engineering, Kansas State University, Manhattan, KS 66506, USA. kejiali@ksu.edu
This study measures pulse wave velocity (PWV) using synchronized wireless pulse oximeters. PWV calculated at inflection points is higher than at waveform feet, reflecting increased arterial pressure.
Area of Science:
- Biomedical Engineering
- Cardiovascular Physiology
- Medical Instrumentation
Background:
- Pulse wave velocity (PWV) is a key indicator of arterial elasticity and cardiovascular health.
- Accurate PWV measurement is crucial for non-invasive assessment of vascular function.
- Existing methods for PWV assessment can be complex or require specialized equipment.
Purpose of the Study:
- To present a novel, accessible method for measuring PWV using readily available wireless pulse oximeters.
- To evaluate the feasibility of calculating PWV from photoplethysmograms (PPGs) acquired from the wrist and fingertip.
- To compare PWV measurements derived from different waveform features (foot, inflection point, peak).
Main Methods:
- Two synchronized wireless pulse oximeters were placed on the wrist and fingertip of the same hand.
- A custom MATLAB interface was developed to acquire and time-align the photoplethysmogram (PPG) signals.
- PWV was calculated by measuring the time difference between PPG features (foot, inflection point, peak) at the two measurement sites.
Main Results:
- PWV values calculated at the inflection point were consistently higher than those calculated at the waveform foot.
- This finding aligns with the physiological understanding that PWV increases with intra-arterial pressure.
- PWV calculations using PPG peaks were complicated by the influence of pulse wave reflections, introducing variable time delays.
Conclusions:
- Synchronized wireless pulse oximeters offer a practical approach for PWV estimation.
- PWV measurement at the inflection point provides a potentially more sensitive indicator of arterial stiffness compared to the foot.
- Care must be taken when interpreting PWV derived from PPG peaks due to the confounding effects of wave reflections.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:51Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of blood pressure in brachial artery(two-step method)