A wavelet based technique for suppression of EMG noise and motion artifact in ambulatory ECG
P Mithun1, Prem C Pandey, Toney Sebastian
1Indian Institute of Technology Bombay, India.
Abstract:
A wavelet-based denoising technique is investigated for suppressing EMG noise and motion artifact in ambulatory ECG. EMG noise is reduced by thresholding the wavelet coefficients using an improved thresholding function combining the features of hard and soft thresholding. Motion artifact is reduced by limiting the wavelet coefficients. Thresholds for both the denoising steps are estimated using the statistics of the noisy signal. Denoising of simulated noisy ECG signals resulted in an average SNR improvement of 11.4 dB, and its application on ambulatory ECG recordings resulted in L(2) norm and max-min based improvement indices close to one. It significantly improved R-peak detection in both the cases.
More Related Videos
11:25Simultaneous Scalp Electroencephalography (EEG), Electromyography (EMG), and Whole-body Segmental Inertial Recording for Multi-modal Neural Decoding
Published on: July 26, 2013
08:08Using Wavelet Entropy to Demonstrate how Mindfulness Practice Increases Coordination between Irregular Cerebral and Cardiac Activities
Published on: May 10, 2017
Related Concept Videos
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...