Related Experiment Video
Updated: May 25, 2026

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
Distributed parameter statics of magnetic catheters
1Stereotaxis, Inc, 4320 Forest Park Ave, Suite 100 St Louis, Missouri 63108, USA. itunay@ieee.org
Abstract:
We discuss how to use special Cosserat rod theory for deriving distributed-parameter static equilibrium equations of magnetic catheters. These medical devices are used for minimally-invasive diagnostic and therapeutic procedures and can be operated remotely or controlled by automated algorithms. The magnetic material can be lumped in rigid segments or distributed in flexible segments. The position vector of the cross-section centroid and quaternion representation of an orthonormal triad are selected as DOF. The strain energy for transversely isotropic, hyperelastic rods is augmented with the mechanical potential energy of the magnetic field and a penalty term to enforce the quaternion unity constraint. Numerical solution is found by 1D finite elements. Material properties of polymer tubes in extension, bending and twist are determined by mechanical and magnetic experiments. Software experiments with commercial FEM software indicate that the computational effort with the proposed method is at least one order of magnitude less than standard 3D FEM.
Related Concept Videos
Magnetic Field Of A Current Loop
Divergence and Curl of Magnetic Field
Magnetic Field Due To A Thin Straight Wire
Magnetostatic Boundary Conditions
Magnetic Field Due to Two Straight Wires
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...