In vitro platelet quality in storage containers used for pediatric transfusions
Sandra Weiss1, Kenneth Scammell, Elena Levin
1Canadian Blood Services Research and Development, Edmonton, Alberta, Canada.
Background:
The in vitro quality of small-volume platelet (PLT) aliquots for pediatric transfusions was assessed to determine the best practice approach.
Study Design And Methods:
Small volumes (50 mL) of single apheresis PLT components (APCs), collected on either CaridianBCT Trima or Haemonetics MCS+ instruments, were aliquoted on Days 2, 3, 4, and 5 postcollection into Fenwal PL1240 or 4R2014 bags or 60-mL polypropylene syringes. Samples were tested for in vitro quality at their recommended expiry times (4 hr for 4R2014 bags and syringes or Day 5 for PL1240 bags). Assays included pH, CD62P expression, and metabolic measures.
Results:
CD62P expression increased throughout storage in all containers. Among the small-volume containers, pH, pCO(2) , lactate, and bicarbonate varied considerably. Regardless of the day of aliquoting, pCO(2) was significantly higher and pO(2) was significantly lower in gas-impermeable syringes than other containers. No bacterial growth was detected in any sample.
Conclusion:
The quality of APCs aliquoted into small-volume containers meets regulatory requirements and is generally equivalent to that of full-volume APCs at expiry.