Comparison of superimposed high-frequency jet ventilation with conventional jet ventilation for laryngeal surgery
R Leiter1, A Aliverti, R Priori
1Anaesthesiology and Intensive Care, Department of Surgical Sciences, Uppsala University, Sjukhusvägen 1, Entrance 70, S-75185 Uppsala, Sweden. robert.leiter@surgsci.uu.se
Background:
New ventilators have simplified the use of supraglottic superimposed high-frequency jet ventilation (SHFJV(SG)), but it has not been systematically compared with other modes of jet ventilation (JV) in humans. We sought to investigate whether SHFJV(SG) would provide more effective ventilation compared with single-frequency JV techniques.
Methods:
A total of 16 patients undergoing minor laryngeal surgery under general anaesthesia were included. In each patient, four different JV techniques were applied in random order for 10-min periods: SHFJV(SG), supraglottic normal frequency (NFJV(SG)), supraglottic high frequency (HFJV(SG)), and infraglottic high-frequency jet ventilation (HFJV(IG)). Chest wall volume variations were continuously measured with opto-electronic plethysmography (OEP), intratracheal pressure was recorded and blood gases were measured.
Results:
Chest wall volumes were normalized to NFJV(SG) end-expiratory level. The increase in end-expiratory chest wall volume (EEV(CW)) was 239 (196) ml during SHFJV(SG) (P<0.05 compared with NFJV(SG)). EEV(CW) was 148 (145) and 44 (106) ml during HFJV(SG) and HFJV(IG), respectively (P<0.05 compared with SHFJV(SG)). Tidal volume (V(T)) during SHFJV(SG) was 269 (149) ml. V(T) was 229 (169) ml (P=1.00 compared with SHFJV(SG)), 145 (50) ml (P<0.05), and 110 (33) ml (P<0.01) during NFJV(SG), HFJV(SG), and HFJV(IG), respectively. Intratracheal pressures corresponded well to changes in both EEV(CW) and V(T). All JV modes resulted in adequate oxygenation. However, was lowest during HFJV(SG) [4.3 (1.3) kPa; P<0.01 compared with SHFJV(SG)].
Conclusion:
SHFJV(SG) was associated with increased EEV(CW) and V(T) compared with the three other investigated JV modes. All four modes provided adequate ventilation and oxygenation, and thus can be used for uncomplicated laryngeal surgery in healthy patients with limited airway obstruction.
Related Concept Videos
Tracheostomy Suctioning II: Procedure
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...