Endovascular retrieval of a dislodged femoral arterial closure device with Alligator forceps
Related Experiment Video
Updated: May 25, 2026

Murine Model of Femoral Artery Wire Injury with Implantation of a Perivascular Drug Delivery Patch
Published on: February 10, 2015
Doeke Boersma1, Marco J van Strijen, Geoffrey T L Kloppenburg
1Department of Vascular Surgery, St Antonius Hospital, Nieuwegein, The Netherlands. d.boersma1@antoniusziekenhuis.nl
The Angio-Seal (St. Jude Medical, Minnetonka, Minn) is a frequently used percutaneous femoral artery closure device to obtain hemostasis after arterial catheterization. Separation and migration of the device, leading to distal embolization and limb-threatening ischemia necessitating emergency surgery, is reported but extremely rare. We present a unique case of successful endovascular bailout management of a dislocated Angio-Seal with use of an Alligator Tooth Retrieval forceps (Cook Medical, London, United Kingdom).
The Angio-Seal (St. Jude Medical, Minnetonka, Minn) is a frequently used percutaneous femoral artery closure device to obtain hemostasis after arterial catheterization. Separation and migration of the device, leading to distal embolization and limb-threatening ischemia necessitating emergency surgery, is reported but extremely rare. We present a unique case of successful endovascular bailout management of a dislocated Angio-Seal with use of an Alligator Tooth Retrieval forceps (Cook Medical, London, United Kingdom).