Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in
X Monnet1, N Anguel, M Jozwiak
1Hôpitaux universitaires Paris-Sud, Hôpital de Bicêtre, service de réanimation médicale, 78, rue du Général Leclerc, Le Kremlin-Bicêtre F-94270, France. xavier.monnet@bct.aphp.fr
Background:
The ability of the third-generation FloTrac/Vigileo software to track changes in cardiac index (CI) induced by volume expansion and norepinephrine in critically ill patients is unknown.
Methods:
In subjects with circulatory failure, we administered volume expansion (20 subjects) and increased (20 subjects) or decreased (20 subjects) the dose of norepinephrine. We measured arterial pressure waveform-derived CI provided by the third-generation FloTrac/Vigileo device (CI(pw)) and transpulmonary thermodilution CI (CI(td)) before and after therapeutic interventions.
Results:
Considering the pairs of measurements performed before and after all therapeutic interventions (n=60), a bias between the absolute values of CI(pw) and CI(td) was 0.26 (0.94) litre min(-1) m(-2) and the percentage error was 54%. Changes in CI(pw) tracked changes in CI(td) induced by volume expansion with moderate accuracy [n=20, bias=-0.11 (0.54) litre min(-1) m(-2), r(2)=0.26, P=0.02]. When changes in CI(td) were induced by norepinephrine (n=40), a bias between CI(pw) and CI(td) was 0.01 (0.41) litre min(-1) m(-2) (r(2)=0.11, P=0.04). The concordance rates between changes in CI(pw) and CI(td) induced by volume expansion and norepinephrine were 73% and 60%, respectively. The bias between changes in CI(pw) and CI(td) significantly correlated with changes in total systemic vascular resistance (r(2)=0.41, P<0.0001).
Conclusions:
The third-generation FloTrac/Vigileo device was moderately reliable for tracking changes in CI induced by volume expansion and poorly reliable for tracking changes in CI induced by norepinephrine.
Related Concept Videos
Heart Failure Drugs: Inotropic Agents
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Heart Failure Drugs: β-Blockers
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...