Accuracy and reliability of length measurements on three-dimensional computed tomography using open-source OsiriX
Gihyeon Kim1, Ho-Joong Jung, Han-Jun Lee
1Department of Radiology, College of Medicine, Chung-Ang University, Seoul, South Korea.
Journal of Digital Imaging
|January 25, 2012
Summary
Three-dimensional computed tomography (3D-CT) with OsiriX software accurately measures bone and joint lengths. This validated method offers high reliability for kinematic analysis, with minimal error compared to direct measurements.
Area of Science:
- Orthopedics
- Medical Imaging
- Biomechanical Engineering
Background:
- Three-dimensional computed tomography (3D-CT) is increasingly used for bone and joint anatomy research.
- Manual processing of CT data for 3D imaging requires validation to minimize systematic and random errors.
Purpose of the Study:
- To evaluate the accuracy and reliability of length measurements on 3D-CT using OsiriX software.
- To compare 3D-CT measurements with gold-standard physical measurements.
Main Methods:
- 14 frozen pig knees underwent 3D-CT scans with five transosseous holes in the femur.
- Lengths between tunnel orifices were measured using digital calipers (gold standard) and OsiriX software.
- Measurements were taken by two observers across two trials to assess reliability and reproducibility.
Main Results:
- Mean differences between OsiriX and caliper measurements were under 0.1 mm, with maximum differences below 0.3 mm.
- No significant differences were found between OsiriX measurements and real measurements (p=0.544 and 0.622).
- Intraclass correlation coefficients (ICC) were very high, indicating excellent agreement between trials and replicates (ICC=0.998 and 0.999).
Conclusions:
- Length measurements on 3D-CT using OsiriX are accurate and highly reliable for kinematic analysis of knees.
- OsiriX software provides a valid alternative to direct physical measurements, with accuracy within 0.3 mm.