Related Experiment Video
Updated: May 25, 2026

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Sudden reversible pacemaker failure in a patient with cardiac sarcoidosis: an unfortunate case of ventricular septal
Nobuhiro Takasugi1, Tomoki Kubota, Itta Kawamura
1Division of Cardiovascular Medicine, Department of Internal Medicine, Gifu University School of Medicine, 1-1 Yanagido, Gifu 501-1194, Japan. konezumi@ccn3.aitai.ne.jp
Abstract:
We report a case of sudden marked deterioration of ventricular stimulation threshold resulting in pacemaker failure 16 months after a ventricular septal lead implantation for atrioventricular block. Echocardiography revealed septal wall thinning at the electrode-tissue interface, which was not detected pre-operatively. Endomyocardial biopsy confirmed cardiac sarcoidosis. The increased threshold was reversible with prednisolone.