Device troubleshooting: cross connection of ventricular leads in a patient with decreased right ventricular
Daisuke Kutsuzawa1, Takanori Arimoto, Tetsuro Shishido
1Department of Cardiology, Pulmonology, and Nephrology, Yamagata University School of Medicine, 2-2-2 Iida-nishi, Yamagata 990-9585, Japan.
Abstract:
A patient with cardiac sarcoidosis showed low R-wave amplitude in the entire right ventricle (RV). To troubleshoot this, a left ventricular (LV) lead was implanted in the coronary vein. The pace/sense terminal of the defibrillation lead was connected to LV:IS-1 of the defibrillator. Conversely, the LV lead was connected to RV:IS-1. Induced ventricular fibrillation was successfully terminated without any undersensing.