Clinical experience with a new endobrochial blocker: the EZ-blocker
Tamás Végh1, Marianna Juhász, Attila Enyedi
1Department of Anesthesiology and Intensive Care, University of Debrecen, Debrecen, Hungary. veghdr@gmail.com
Purpose:
In some thoracic surgical procedures, the insertion of a double-lumen tube (DLT) is not feasible, or the altered use of a DLT and a single-lumen tube (SLT) is justified during the surgery. In the present article we report our experience with a new bronchial blocker, the EZ-blocker, in clinical use.
Methods:
Data were obtained from ten patients undergoing thoracic surgery necessitating one-lung ventilation. For lung isolation, a combination of an SLT and an EZ-blocker was used. The time of insertion and positioning of the EZ-blocker, the lung deflation time with the EZ-blocker cuff inflated and deflated, and the cuff's minimal occlusion volume were recorded. Based on the CT scan, the diameter of the main bronchi and the angle of the bifurcation were measured offline.
Results:
The insertion duration of the EZ-blocker was 76 ± 15 s. Two malpositionings were caused by the too-deep positioning of the SLT used for introducing the EZ-blocker, which could be corrected within 65 ± 7 s. The use of the EZ-blocker allowed a short deflation time of the lung without (9.4 ± 0.7 s) and with (4.1 ± 0.7 s) administration of suctioning. The proper block was only dependent on the diameter of the main bronchi and was independent of the bifurcation angle.
Conclusions:
Use of the EZ-blocker is easy and safe. The short insertion time and short lung deflation time through the lumen of the SLT allows its use in emergency situations or in cases of a difficult airway.
Related Concept Videos
Treatment for Pulmonary Arterial Hypertension: Endothelin Receptor Antagonists
ETs are synthesized through a complex sequence of enzymatic steps, primarily involving an enzyme referred to as endothelin-converting enzyme (ECE). Of...
COPD: Management Using Bronchodilators and Corticosteroids
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Antiasthma Drugs: β2-Adrenoceptor Agonists
One class of bronchodilators includes β2-adrenoceptor agonists. These agents target the β2-adrenoceptors located on bronchial smooth muscle cells. By stimulating these receptors, β2-agonists induce relaxation in these...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Inhaled Medications