Related Experiment Video
Updated: May 25, 2026

07:06
Destabilization of the Medial Meniscus and Cartilage Scratch Murine Model of Accelerated Osteoarthritis
Published on: July 6, 2022
Tear meniscus measurement by spectral optical coherence tomography
Grzegorz Czajkowski1, Bartlomiej J Kaluzny, Adriana Laudencka
1Department of Ophthalmology, Collegium Medicum, Nicolaus Copernicus University, Bydgoszcz, Poland. gcz5@wp.pl
Summary
Spectral Optical Coherence Tomography (OCT) measurements of lower tear meniscus parameters, specifically tear meniscus area (TMA) and tear meniscus height (TMH), show high accuracy in diagnosing dry eye syndrome. These OCT findings correlate well with traditional dry eye tests and patient-reported symptoms.
Area of Science:
- Ophthalmology
- Medical Imaging
- Dry Eye Disease Research
Background:
- Dry eye syndrome is a prevalent condition affecting ocular surface health.
- Accurate diagnosis relies on various clinical tests, including Schirmer test and tear break-up time.
- Quantitative tear evaluation methods are crucial for objective diagnosis.
Purpose of the Study:
- To assess the utility of lower tear meniscus measurements using anterior segment Spectral Optical Coherence Tomography (OCT).
- To determine the diagnostic accuracy (sensitivity and specificity) of different tear meniscus parameters for dry eye syndrome.
- To correlate OCT-derived tear meniscus measurements with established clinical tests and patient-reported symptoms.
Main Methods:
- 111 eyes from 111 patients underwent Spectral OCT for tear meniscus evaluation.
- Measurements included tear meniscus cross-section area (TMA), tear meniscus height (TMH), and tear meniscus depth (TMD).
- Standard ophthalmic examination, Ocular Surface Disease Index (OSDI) questionnaire, tear break-up time, and Schirmer tests were performed.
Main Results:
- TMA and TMH showed the highest correlation with Schirmer test results (Spearman's rho = 0.54 and 0.52, respectively).
- TMA, TMH, and TMD measurements were significantly lower in patients with dry eye compared to controls.
- Tear meniscus height (TMH) demonstrated 80.56% sensitivity and 89.33% specificity; TMA showed 86.11% sensitivity and 85.33% specificity for dry eye diagnosis.
Conclusions:
- Lower tear meniscus parameters measured by Spectral OCT correlate well with the Schirmer test, tear break-up time, and subjective dry eye symptoms.
- Tear meniscus area (TMA) and tear meniscus height (TMH) offer high sensitivity and specificity for diagnosing dry eye syndrome.
- Spectral OCT provides a valuable quantitative tool for evaluating tear film and diagnosing dry eye disease.