Alternative X-ray filters for an intra-oral digital radiographic system
J Stecke1, A D Cruz, S M Almeida
1Piracicaba Dental School of the State University of Campinas, Sao Paulo, Brazil.
Objective:
The aim of this study was to evaluate the effect of the modulation of the radiation spectrum with the use of alternative X-ray filters in the quality of intra-oral digital images from storage phosphor plates.
Methods:
The radiographic exposures were performed in a GE 1000 X-ray machine (General Electric Co., Milwaukee, WI), operating at 65 kVp, 10 mA, 40 cm focus receptor distance using three different exposure times: 0.05 s, 0.16 s and 0.35 s. The control filter (GC) was 100% aluminium (Al) with a thickness of 1.5 mm. The tested filters were: G1, 97% Al and 3% copper (Cu) with 1.47 mm thickness; G2, 96% Al and 4% Cu with 1.53 mm thickness; G3, 95% Al and 5% zinc (Zn) with 1.56 mm thickness; G4, 98% Al and 2% Zn with 1.5 mm thickness; and G5, 95% Cu and 5% Zn with 1.6 mm thickness. For formation of the image, a 12-step Al wedge (each step with increments of 1 mm in thickness) was radiographed 10 times. Pixel values measured in digital images were converted into optical density (OD).
Results:
All replicates showed OD with high reproducibility (r > 0.95) for all exposure times and tested filters. In comparison between filters, statistically significant differences in density (p < 0.05) were observed. The OD curve of the G5 filter in all exposure times and G3 filter in an exposure time of 0.05 s showed changes in shape (p < 0.05).
Conclusions:
Excluding the G5 filter, all others tested filters can be used as a substitute for GC without losses in image quality.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Determination of Crystal Structures