The Intra-Aortic Balloon Pump
Published on: February 5, 2021
Pressure-wave energy relationship during IABP counterpulsation in a mock circulation: changes with angle and
Giovanni Biglino1, Christina Kolyva, Ashraf W Khir
1Brunel Institute for Bioengineering, Brunel University, Uxbridge, UK.
Purpose:
Despite decades of successful clinical use of the intra aortic balloon pump (IABP), certain aspects of its operation are not yet fully understood. This work aims to investigate in vitro the mechanism underlying balloon inflation and deflation with varying assisting frequency and operating angle with respect to the horizontal, by studying the corresponding pressure and wave energy changes.
Methods:
A mock circulatory system (MCS), with physiological distribution of peripheral resistance and compliance, presented a controllable test bed. We used Wave Intensity Analysis (WIA) to identify balloon-generated waves and quantify their energy. Conventional hemodynamic parameters were also calculated. Tests were repeated at varying operating angles (0°-45°), resembling the semi-recumbent position in the ICU, and at different assisting frequencies (1:1, 1:2, 1:3). Two balloons (25 cc and 40 cc in volume) were tested.
Results:
The main waves associated with counterpulsation were identified as a backward compression wave associated with balloon inflation and a backward expansion wave associated with balloon deflation. Results showed that the IABP inflation and deflation benefits are reduced with increasing angle, in terms of the size of the inflation and deflation waves as well as in terms of diastolic pressure augmentation and end-diastolic pressure reduction. Both WIA findings and pressure parameters indicated 1:1 as the most effective mode of pumping.
Conclusions:
This study shows that, in vitro, a greater benefit of counterpulsation can be achieved in the horizontal position at 1:1 assisting frequency, with a good correlation between wave and pressure results.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical indicator...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation