[Determination of metal by liquid electrode discharge atomic emission spectrometry with a new design]
Ya-qin Yang1, Hong-mei Wang, Yan-nan Chu
1Laboratory of Environment Spectroscopy, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China yulingin@hotmail.com
Abstract:
A liquid electrode discharge atomic emission spectrometry has been developed with a new design. Plasma fluctuation from the variations in the gap between the W anode and liquid cathode was eliminated by providing a sawtooth-shaped drainage structure. The sawtooth was placed at a distance of 2 mm from the glass capillary top. When a solution is pumped through a glass capillary, the liquid drop forms and rises and on reaching the maximum size collapses while the next drop forms. This results in a continuous change in the interelectrode distance which will cause the discharge current to change, plasma flickers and is extinguished. With the new design, the surface tension of the drop was destroyed before its growing up, and the solution was flowing along the drainage groove. The discharge can last for 3 hours at least with such arrangement. The analytical response curves for lead, chromium, cadmium, zinc, vanadium, nickel, copper, silver and cobalt demonstrated good linearity. The limit of detections of lead, chromium, cadmium, zinc, vanadium nickel, copper, silver and cobalt were determined to be 0.08, 0.61, 0.48, 1.60, 10.88, 0.48, 0.084, 0.048 and 0.27 mg x L(-1) respectively.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Overview
Electrodeposition
Electrodeposition can...
Electrogravimetric Analysis: Overview
To test the completeness of the...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.