Exploring the possibilities of cryogenic cooling in liquid chromatography for biological applications: a proof of
Hamed Eghbali1, Koen Sandra, Bart Tienpont
1Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, B-1050 Brussels, Belgium. heghbali@gmail.com
Abstract:
The possibilities to use cryogenic cooling to trap components in liquid chromatography was investigated. In a first step, van 't Hoff plots were measured with a reversed-phase column using the temperature control unit of a conventional high performance liquid chromatography (HPLC) system to gain insight in the retention behavior of proteins at low temperatures. It was estimated that retention factors in the range of k = 10(4) could be achieved at T = -20 °C for lysozyme, indicating that temperature is a usable parameter to trap components in LC. In a next step, trapping experiments were carried out on a nano-LC system, equipped with a UV-detector, using a commercial reversed-phase column. An in-house built setup, allowing cooling of a segment of the column down to temperatures below T = -20 °C, was used to trap components. Experiments were conducted under isocratic and gradient conditions with methanol as organic solvent. It is demonstrated that, by thermally trapping and elution of components, an enhanced S/N ratio and decreased peak widths can be obtained. At the same time, a significant increase in pressure drop occurs during the cooling process. Limitations and benefits of the technique are further discussed.
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Instrumentation
Cryo-electron Microscopy