Comparison of whole blood fibrin-based clot tests in thrombelastography and thromboelastometry
Cristina Solomon1, Benny Sørensen, Gerald Hochleitner
1Department of Anaesthesiology and Intensive Care, Salzburger Landeskliniken SALK, 48 Müllner Hauptstrasse, 5020 Salzburg, Austria. Solomon.Cristina@googlemail.com
Background:
Fibrin-based clot firmness is measured as maximum amplitude (MA) in the functional fibrinogen (FF) thrombelastographic assay and maximum clot firmness (MCF) in the FIBTEM thromboelastometric assay. Differences between the assays/devices may be clinically significant. Our objective was to compare clot firmness parameters through standard (FF on a thrombelastography device [TEG®]; FIBTEM on a thromboelastometry device [ROTEM®]) and crossover (FF on ROTEM®; FIBTEM on TEG®) analyses.
Methods:
Whole-blood samples from healthy volunteers were subjected to thrombelastography and thromboelastometry analyses. Samples were investigated native and following stepwise dilution with sodium chloride solution (20%, 40%, and 60% dilution). Samples were also assessed after in vitro addition of medications (heparin, protamine, tranexamic acid) and 50% dilution with hydroxyethyl starch, gelatin, sodium chloride, and albumin.
Results:
FF produced higher values than FIBTEM, regardless of the device, and TEG® produced higher values than ROTEM®, regardless of the assay. With all added medications except heparin 400 U/kg bodyweight, FF MA remained significantly higher (P < 0.05) than FIBTEM MCF, which was largely unchanged. FF MA was significantly reduced (P = 0.04) by high-dose heparin and partially restored with protamine. Fifty percent dilution with hydroxyethyl starch, albumin, and gelatin decreased FIBTEM MCF and FF MA by >50%.
Conclusions:
These results demonstrate differences when measuring fibrin-based clotting via the FF and FIBTEM assays on the TEG® and ROTEM® devices. Point-of-care targeted correction of fibrin-based clotting may be influenced by the assay and device used. For the FF assay, data are lacking.
More Related Videos
06:16Tracking Fibrinolysis of Chandler Loop-Formed Whole Blood Clots Under Shear Flow in An In-Vitro Thrombolysis Model
Published on: April 19, 2024
08:01The Nijmegen Hemostasis Assay: Simultaneous Fluorogenic Measurement of Thrombin and Plasmin Generation in a Single Well
Published on: February 27, 2026
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies