Eye doses to staff in a nuclear medicine department
Elizabeth C Summers1, Janis L E Brown, Peter J Bownes
1Medical Physics Department, Mount Vernon Hospital, Northwood, Middlesex, UK.
Objective:
Occupational radiation doses to the Nuclear Medicine Department staff at Mount Vernon Hospital are routinely measured using optically stimulated luminescence dosemeters for whole-body effective dose and ring thermoluminescence dosemeters (TLDs) for finger dose. In 2002, a project was carried out using LiF:Mg,Cu,P Chinese TLDs to measure the dose to the lens of the eye received by staff during normal working procedures.
Methods:
Separate pairs of TLDs were worn by staff on their forehead between their eyes while dispensing and releasing in the radiopharmacy, injecting, and when administering I-131 capsules to patients. The dose received was calculated using calibration data from identical TLDs irradiated with Tc-99m, I-131, and the Ir-192 source of a Gammamed High Dose Rate (HDR) treatment unit. Data were collected over a 5-month period and the mean dose to the eye was calculated for each procedure.
Results:
Using a typical yearly workload, the annual dose to the eye for a single member of staff was calculated and found to be 4.5 mSv.
Conclusion:
The occupational eye dose limit was, at the time, 150 mSv; therefore, staff were well below the level (3/10th of this limit) that would have required them to be classified. However, there have been large increases in radiopharmacy production and I-131 therapies administered at Mount Vernon in subsequent years. It is therefore expected that the eye dose received by staff will have increased to be significantly higher than 4.5 mSv and will in fact be greater than 6 mSv, which is 3/10th of the proposed new dose limit and would require these staff to become classified workers.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Biological Effects of Radiation
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET