Development and validation of rapid use scope test strips to determine the efficacy of manual cleaning for flexible

Michelle J Alfa1, Nancy Olson, Pat Degagné

  • 1Diagnostic Services of Manitoba, Winnipeg, Manitoba, Canada; St Boniface Research Centre, Winnipeg, Manitoba, Canada. malfa@dsmanitoba.ca

Abstract

Insights

The Rapid Use Scope Test (RUST) effectively assesses flexible endoscope cleaning. While most channels passed, specific areas like elevator guide-wire and endoscopic retrograde cholangiopancreatography channels require improved manual cleaning protocols.

Area of Science:

  • Endoscopy
  • Medical Device Reprocessing
  • Infection Control

Background:

  • Manual cleaning of flexible endoscopes is frequently inadequate.
  • A need exists for reliable methods to assess cleaning efficacy.

Purpose of the Study:

  • To validate the sample collection protocol and the Rapid Use Scope Test (RUST).
  • To assess the clinical utility of the RUST test for evaluating endoscope cleanliness.

Main Methods:

  • Established cleaning benchmarks for protein, hemoglobin, and carbohydrate levels.
  • Validated the RUST audit tool through simulated and in-use testing in 43 Canadian endoscopy clinics.

Main Results:

  • Simulated testing confirmed RUST's ability to flag inadequately cleaned endoscopes.
  • Clinical use showed 96.6% of channels were RUST negative, but specific channels exceeded benchmarks.
  • Reprocessing staff recognized the value of RUST for compliance assessment.

Conclusions:

  • The validated RUST test serves as a rapid audit tool for manual endoscope cleaning.
  • RUST can be integrated into endoscopy quality assurance programs to improve patient safety.