Development and validation of rapid use scope test strips to determine the efficacy of manual cleaning for flexible
Michelle J Alfa1, Nancy Olson, Pat Degagné
1Diagnostic Services of Manitoba, Winnipeg, Manitoba, Canada; St Boniface Research Centre, Winnipeg, Manitoba, Canada. malfa@dsmanitoba.ca
Background:
Cleaning of flexible endoscopes is most commonly performed using manual methods that are often performed inadequately. The aim of this study was to validate the sample collection protocol and the Rapid Use Scope Test (RUST) and then assess its usefulness in clinical use.
Methods:
The benchmarks for adequate cleaning were protein <6.4 μg/cm(2), hemoglobin <2.2 μg/cm(2), and carbohydrate <1.2 μg/cm(2). Sample collection consisted of flushing 10 mL of sterile reverse osmosis water through the suction-biopsy port to the distal end. Validation of the RUST audit tool included simulated-use and in-use testing in 43 endoscopy clinics across Canada.
Results:
Simulated-use testing validated that improperly cleaned endoscopes that exceeded the cleaning benchmarks would be flagged by the RUST test. The clinical-use study indicated that 96.6% of 1,489 scope channels tested were RUST negative; however, 19% and 12% of elevator guide-wire channels and endoscopic retrograde colangiopancreatography channels, respectively, exceeded the benchmarks. The survey indicated that reprocessing personnel valued a rapid audit tool for assessing compliance with manual cleaning.
Conclusion:
The validated RUST test provides health care users with a rapid audit tool for manual cleaning that can be integrated into the quality program in endoscopy.
Insights
The Rapid Use Scope Test (RUST) effectively assesses flexible endoscope cleaning. While most channels passed, specific areas like elevator guide-wire and endoscopic retrograde cholangiopancreatography channels require improved manual cleaning protocols.
Area of Science:
- Endoscopy
- Medical Device Reprocessing
- Infection Control
Background:
- Manual cleaning of flexible endoscopes is frequently inadequate.
- A need exists for reliable methods to assess cleaning efficacy.
Purpose of the Study:
- To validate the sample collection protocol and the Rapid Use Scope Test (RUST).
- To assess the clinical utility of the RUST test for evaluating endoscope cleanliness.
Main Methods:
- Established cleaning benchmarks for protein, hemoglobin, and carbohydrate levels.
- Validated the RUST audit tool through simulated and in-use testing in 43 Canadian endoscopy clinics.
Main Results:
- Simulated testing confirmed RUST's ability to flag inadequately cleaned endoscopes.
- Clinical use showed 96.6% of channels were RUST negative, but specific channels exceeded benchmarks.
- Reprocessing staff recognized the value of RUST for compliance assessment.
Conclusions:
- The validated RUST test serves as a rapid audit tool for manual endoscope cleaning.
- RUST can be integrated into endoscopy quality assurance programs to improve patient safety.