Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to serve as the standard temperature of the blackbody. The errors of measurement for the two meters were reduced by the calibration equation. The predictive values could meet the requirements of the ASTM standard. The sources of uncertainty include the standard deviations of replication at fixed temperature or the predicted values of calibration equation, reference standard values and resolution. The uncertainty analysis shows that the uncertainty of calibration equation is the main source for combined uncertainty. Ambient temperature did not have the significant effects on the measured performance. The calibration equations could improve the accuracy of ITTs. However, these equations did not improve the uncertainty of ITTs.
More Related Videos
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...