Related Experiment Video
Updated: May 25, 2026

06:09
Piezo High Accuracy Surgical Osteal Removal (PHASOR): A Technique for Improved Cranial Window Surgery in Mice
Published on: March 2, 2018
Stroke dynamics and frequency of 3 phacoemulsification machines
Daniele Tognetto1, Paolo Cecchini, Pia Leon
1University Eye Clinic of Trieste, Trieste, Department of Biomedical Science, University of Chieti, Pescara, Italy. tognetto@units.it
Journal of Cataract and Refractive Surgery
|February 11, 2012
Summary
This study measured phacoemulsification machine performance, revealing significant differences in working frequency and stroke dynamics among three models. Understanding these unique tip movements can optimize cataract surgery lens removal.
Area of Science:
- Ophthalmology
- Surgical Technology
- Biomedical Engineering
Background:
- Phacoemulsification is a key cataract surgery technique.
- Understanding phaco tip dynamics is crucial for surgical efficiency and patient outcomes.
Purpose of the Study:
- To quantify and compare the working frequency and stroke dynamics of three phacoemulsification machines.
- To analyze tip displacement and movement patterns under varying power settings.
Main Methods:
- An experimental study utilizing video wet fixtures with high-speed cameras and strobe lighting.
- Measurement of working frequency and vectorial tip displacement at different phacoemulsification powers.
Main Results:
- Significant variations in working frequency (28.8-43.0 kHz) and stroke dynamics were observed across the Infiniti Ozil, Whitestar Signature, and Stellaris machines.
- The Stellaris machine exhibited the largest longitudinal stroke (139 μm).
- Distinct lateral and elliptical movement patterns were noted for the Infiniti Ozil and Whitestar Signature machines, with specific characteristics in torsional and Ellips FX handpiece modes.
Conclusions:
- The three phacoemulsification machines demonstrate distinct performance profiles regarding frequency and stroke dynamics.
- Surgeons can leverage knowledge of specific machine stroke characteristics (e.g., lateral, elliptical) to enhance lens removal efficacy.