Related Experiment Video
Updated: May 25, 2026

A Bedside, Single Burr Hole Approach to Multimodality Monitoring in Severe Brain Injury
Published on: March 26, 2019
Implantable ICP monitor for improved hydrocephalus management
Ellyce Stehlin1, Simon Malpas, Peter Heppner
1Auckland Bioengineering Institute, University of Auckland, Auckland, New Zealand.
Background:
Hydrocephalus patients are commonly treated by insertion of ventriculoperitoneal shunts, but these have high complication rates. Monitoring of shunt and patient condition can be achieved through measuring intracranial pressure (ICP). Significant zero drift has limited previous developments towards a long-term implantable ICP monitor. We present a new implantable solid-state pressure sensor system appropriate for chronic (lifetime) monitoring of ICP.
Materials And Methods:
Initial designs of the proposed ICP system were realised and the pressure sensor catheter underwent bench-top tests to analyse its characteristics. A drift rig was constructed for the long-term analysis of the sensor's zero drift. The pressure sensor catheter was used to continuously monitor blood pressure in rats.
Results:
Three potential design solutions were realised: a standalone sensor, the sensor unit in line with a shunt system, and the sensor unit fully integrated into the shunt valve housing. Initial stability results across 46 days show a maximum drift of less than 2 mmHg and a minimum drift of less than 0.2 mmHg.
Conclusion:
Initial experience with the new implantable solid-state pressure sensor system confirms its suitability for chronic pressure monitoring. The device is promising for providing vital information on shunt and patient condition.
Related Concept Videos
Increased Intracranial Pressure l: Introduction
Increased Intracranial Pressure ll: Pathophysiology