On the method of a randomised comparison of programmable valves with and without gravitational units: the SVASONA
Johannes Lemcke1, Ullrich Meier, Cornelia Müller
1Department of Neurosurgery, Unfallkrankenhaus, Berlin, Germany. johannes.lemcke@ukb.de
Objective:
The supremacy of low-pressure valves (LPV) in the therapy of patients with idiopathic normal pressure hydrocephalus (iNPH) has been proven by the Dutch NPH study. The downside of LPVs is the high rate of overdrainage complications. In the meantime gravitational units have been developed with the objective of minimising overdrainage complications. Do these gravitational units allow the same favourable outcomes as in the Dutch NPH study without overdrainage complications? The goal of this prospective randomised controlled multicentre trial is to compare the rate of overdrainage complications after shunt surgery with programmable valves with or without a gravitational unit.
Method:
Patients suspected with iNPH with gait ataxia and one or two other symptoms of Hakim's triad and enlarged ventricles on CT or MRI underwent a cerebrospinal fluid (CSF) infusion test, CSF tap test and/or continuous ICP measurement in seven experienced centres. If iNPH was diagnosed the patients were randomised to receive a ventriculoperitoneal shunt with a programmable valve with or without a gravitational unit. The patients were followed up for 3, 6 and 12 months after surgery. The primary hypothesis of the study was that programmable valves with gravitational units will reduce the rate of overdrainage from 25% to 10%. The outcome measured by iNPH-specific outcome scales (Kiefer score, Black grading scale) and the generic quality of life (short form 12, SF12) was defined as the secondary endpoint. To detect the hypothesised difference in the incidence of overdrainage with a type I error of 5% and a type II error of 20%, correcting for multiple testing and an anticipated drop-out rate of 10%, we planned to enrol 250 patients. To ensure safety and efficacy we performed a planned interim analysis halfway.
Results:
The study design has been found to be effective with regard to the primary hypothesis.
Conclusion:
The design of the SVASONA study was developed to be able to confirm the primary hypothesis. Thus, the method of the study should solve the dilemma of the Dutch NPH study by the randomised comparison of LPVs with and without gravitational units.
Related Concept Videos
Typical Model Studies
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Gravimetry: Overview
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Frequency of Spring-Mass System
Consider a block on a spring on a frictionless surface. There...
Variation of Atmospheric Pressure
Assuming the air temperature is constant at a given altitude and that the ideal gas law of thermodynamics describes the atmosphere to a good approximation, one can find the variation of atmospheric pressure with height.
Let p(y) be the atmospheric pressure at...