Rapid radiofrequency calibration in MRI.

J W Carlson1, D M Kramer

  • 1Radiologic Imaging Laboratory, University of California, San Francisco 94080.

Magnetic Resonance in Medicine
|September 1, 1990
PubMed
Summary

This study introduces new methods for setting radiofrequency (rf) transmitter levels using a three-pulse sequence. The techniques rapidly estimate the flip angle to prevent systematic errors from relaxation during rf pulses.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The complex polyploid genome architecture of sugarcane.

Nature·2024
Same author

Combination of Proactive Molecular Risk Classifier for Endometrial cancer (ProMisE) with sonographic and demographic characteristics in preoperative prediction of recurrence or progression of endometrial cancer.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2020
Same author

Who cares? The impact on caregivers of suspected mining-related lung cancer.

Current oncology (Toronto, Ont.)·2019
Same author

Validation of ultrasound strategies to assess tumor extension and to predict high-risk endometrial cancer in women from the prospective IETA (International Endometrial Tumor Analysis)-4 cohort.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2019
Same author

Ultrasound characteristics of endometrial cancer as defined by International Endometrial Tumor Analysis (IETA) consensus nomenclature: prospective multicenter study.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2017
Same author

A portable multi-flash kinetic fluorimeter for measurement of donor and acceptor reactions of Photosystem 2 in leaves of intact plants under field conditions.

Photosynthesis research·2014

Area of Science:

  • Physics
  • Magnetic Resonance Imaging

Background:

  • Accurate radiofrequency (rf) transmitter calibration is crucial for quantitative magnetic resonance imaging (MRI).
  • Traditional methods can be time-consuming and susceptible to errors caused by relaxation effects during pulse sequences.

Purpose of the Study:

  • To present modified procedures for rapid and accurate setting of rf transmitter levels.
  • To develop a method that minimizes systematic errors caused by relaxation during the pulse sequence.

Main Methods:

  • Utilizes a three-pulse sequence for rf level calibration.
  • Calculates an estimate of the flip angle based on a geometrically weighted ratio of four acquired signals.
  • Employs an updating prescription for iterative adjustment of rf levels.

Main Results:

  • The described procedures allow for rapid estimation of the flip angle.
  • The method effectively eschews systematic errors arising from relaxation during the pulse sequence.
  • Provides a robust approach for calibrating rf transmitter output.

Conclusions:

  • The modified procedures offer a fast and accurate solution for rf transmitter level setting in MRI.
  • This technique improves quantitative accuracy by mitigating relaxation-induced errors.
  • The method is suitable for applications requiring precise control over rf pulse amplitudes.