[A biomechanical study on internal fixation of proximal ulna combined with olecranon fracture]
Lin Teng1, Gang Zhong, Gang Liu
1Department of Orthopedics, West China Hospital, Sichuan University, Chengdu Sichuan, 610041, P R China.
Objective:
To compare the biomechanical stability of Kirschner wire and tension band wiring, reconstruction plate combined with tension band wiring, and olecranon anatomical plate in fixing proximal ulna combined with olecranon fracture, so as to provide the theoretical evidence for clinical selection of internal fixation.
Methods:
Eight specimens of elbow joints and ligaments were taken from eight fresh male adult cadaveric elbows (aged 26-43 years, mean 34.8 years) donated voluntarily. The model of proximal ulna combined with olecranon fracture was made by an osteotomy in each specimen. Fracture end was fixed successively by Kirschner wire and tension band wiring (group A), reconstruction plate combined with tension band wiring (group B), and olecranon anatomical plate (group C), respectively. The biomechanical test was performed for monopodium compression experiments, and load-displacement curves were obtained. The stability of the fixation was evaluated according to the load value when the compression displacement of fracture segment was 2 mm.
Results:
No Kirschner wire withdrawal, broken plate and screw, loosening and specimens destruction were observed. The load-displacement curves of 3 groups showed that the displacement increased gradually with increasing load, while the curve slope of groups B and C was significantly higher than that of group A. When the compression displacement was 2 mm, the load values of groups A, B, and C were (218.6 +/- 66.9), (560.3 +/- 116.1), and (577.2 +/- 137.6) N, respectively; the load values of groups B and C were significantly higher than that of group A (P < 0.05), but no significant difference was observed between groups B and C (t = 0.305, P = 0.763).
Conclusion:
The proximal ulna combined with olecranon fracture is unstable. Reconstruction plate combined with tension band wiring and olecranon anatomical plate can meet the requirement of fracture fixation, so they are favorable options for proximal ulna combined with olecranon fracture. Kirschner wire and tension band wiring is not a stable fixation, therefore, it should not be only used for proximal ulna combined with olecranon fracture.
Related Concept Videos
Bones of the Upper Limb: Ulna
Bones of the Upper Limb: Humerus
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...