Determination of propofol using high performance liquid chromatography in whole blood with fluorescence detection
Jason Yarbrough1, Ralph Harvey, Sherry Cox
1University of Tennessee, Departments of Biomedical and Diagnostic Sciences, College of Veterinary Medicine, Knoxville, TN 37996, USA. jyarbro2@utk.edu
Abstract:
High-performance liquid chromatography method for the determination of propofol has been developed and validated. Following a liquid extraction using ethyl acetate and hexane, samples were separated by reverse-phase high-performance liquid chromatography on an XBridge C(18) column and quantified using fluorescence detection at an excitation of 276 nm and an emission of 310 nm. The mobile phase was a mixture of water (pH 4.0) and acetonitrile, with a flow rate of 1.5 mL/min. The standard curve ranged from 5-2000 ng/mL. Intra- and inter-assay variability for propofol was less than 10%, and the average recovery was greater than 95%. This assay is suitable for use in pharmacokinetic studies.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: