Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Influence of cone-beam computed tomography enhancement filters on diagnosis of simulated external root resorption
Sergio Lins de Azevedo Vaz1, Taruska Ventorini Vasconcelos, Frederico Sampaio Neves
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, State University of Campinas, Piracicaba, São Paulo, Brazil. sergiolinsv@gmail.com
Introduction:
The aim of this study was to determine if cone-beam computed tomography (CBCT) enhancement filters influence the diagnosis of simulated external root resorption (ERR).
Methods:
Buccal, mesial, and distal root surfaces of 20 single-rooted premolar teeth were evaluated for the presence or absence of simulated defects. Images were obtained using a Classic i-CAT CBCT device (Imaging Sciences International, Inc, Hatfield, PA) and analyzed by 3 oral radiologists. Evaluations were performed with and without CBCT filters. Accuracy, sensitivity, specificity, and positive and negative predictive values were determined. The McNemar test verified the disagreement between all images versus the gold standard and original images versus images with filters; P values < .05 were considered statistically significant. Inter- and intraobserver agreement was calculated by the kappa test.
Results:
The means of intra- and interobserver agreement ranged from good to excellent. Sharpen 3×3 had the highest sensitivity (0.64), accuracy (0.62), and negative predictive value (0.33). Specificity was the highest (0.56) for S9, Smooth, Smooth 3×3, and Angio Sharpen High filters. Positive predictive values were the highest (0.81) for the Smooth 3×3 and Angio Sharpen High filters. The McNemar test showed statistically significant differences from the gold standard for all images (P < .05) as well as for original images versus images with the Shadow filter (P = .01). The other filters did not show statistically significant differences from the original images.
Conclusions:
The influence of enhancement CBCT filters on ERR diagnosis is small. The Sharpen 3×3 filter gave the best results and the Shadow filter the worst results.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...