Relative accuracy of arterial and capillary glucose meter measurements in critically ill patients
L Lonjaret1, V Claverie, E Berard
1Department of Anesthesiology and Intensive Care, University Paul-Sabatier, University Hospital of Toulouse, Place du Dr-Baylac, 31059 Toulouse cedex 9, France. lonjaret.l@chu-toulouse.fr
Aim:
As optimizing glucose control in critically ill patients remains a challenge for intensive-care physicians, this study aimed to determine the accuracy of glucose measurements.
Methods:
The accuracy of capillary and arterial blood glucose meter measurements was compared with central laboratory arterial glucose measurements; the factors associated with inaccurate measures were also determined.
Results:
Altogether, 302 samples from 75 patients were assessed. Mean glucose levels were 126±52 mg/dL for capillary measurements, 133±50 mg/dL for arterial measurements and 143±54 mg/dL for serum glucose laboratory measurements. Compliance with the ISO 15197 guidelines was observed in 74.8% of the capillary samples and 88.7% of the arterial samples. However, all measurements by glucose meter (with either capillary or arterial samples) led to underestimations of serum glucose.
Conclusion:
In critically ill patients, glucose measurements from capillary and arterial blood by glucose meter are inaccurate, and can potentially lead to inappropriate use of insulin-infusion protocols and failure to achieve glycaemic targets.