Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
Comparison between three-dimensional speckle-tracking echocardiography and cardiac magnetic resonance imaging for
Sebastiaan A Kleijn1, Wessel P Brouwer, Mohamed F A Aly
1Department of Cardiology 5F 003, Institute for Cardiovascular Research (ICaR-VU), VU University Medical Center, De Boelelaan 1117, 1081 HV Amsterdam, The Netherlands. s.kleijn@vumc.nl
Aims:
We evaluated the accuracy of three-dimensional speckle-tracking echocardiography (3DSTE) to evaluate left ventricular (LV) volumes, ejection fraction (EF), and global circumferential strain (CS) in comparison with cardiac magnetic resonance imaging (MRI) in a healthy population.
Methods And Results:
A total of 45 out of 50 consecutive healthy subjects (38 males, age 45 ± 15 years) successfully underwent both 3DSTE and MRI on the same day. Three-dimensional echocardiography data sets were analysed using speckle tracking to measure LV end-diastolic and end-systolic volumes, EF, and global CS. With MRI, the method of discs approximation was used to obtain volumes and the EF, whereas CS was acquired using myocardial tissue tagging. Inter-technique comparisons included regression and the Bland-Altman analysis. For quantification of LV volumes, 3DSTE correlated well with MRI (r: 0.75-0.81), but volumes were significantly underestimated with relatively large biases (13-34 mL) and wide limits of agreement (SD: 11-25 mL). However, excellent accuracy was revealed for measurement of EF by 3DSTE with a good correlation (r: 0.91), minimal bias, and narrow limits of agreement (0.6 ± 1.7%) compared with MRI. For measurement of CS, a large mean bias was found between techniques (10.0%), despite narrow limits of agreement (SD: 1.7%) and a good correlation between techniques (r: 0.80).
Conclusion:
Although 3DSTE-derived LV volumes are underestimated in most patients compared with MRI, measurement of the LVEF revealed excellent accuracy. Measurements of CS were systematically greater (i.e. more negative) with 3DSTE than MRI, which likely reflects various inter-technique differences that preclude direct comparability of their measurements.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...