Related Experiment Video
Updated: May 24, 2026

Automated System for Single Molecule Fluorescence Measurements of Surface-immobilized Biomolecules
Published on: November 2, 2009
Line-monitoring, hyperspectral fluorescence setup for simultaneous multi-analyte biosensing
1Laboratory of Optical Imaging and Sensing, Graduate School at Shenzhen, Tsinghua University, Shenzhen 518055, China. liuzhiyi07@mails.tsinghua.edu.cn
Abstract:
Conventional fluorescence scanners utilize multiple filters to distinguish different fluorescent labels, and problems arise because of this filter-based mechanism. In this work we propose a line-monitoring, hyperspectral fluorescence technique which is designed and optimized for applications in multi-channel microfluidic systems. In contrast to the filter-based mechanism, which only records fluorescent intensities, the hyperspectral technique records the full spectrum for every point on the sample plane. Multivariate data exploitation is then applied to spectra analysis to determine ratios of different fluorescent labels and eliminate unwanted artifacts. This sensor is designed to monitor multiple fluidic channels simultaneously, providing the potential for multi-analyte biosensing. The detection sensitivity is approximately 0.81 fluors/μm2, and this sensor is proved to act with a good homogeneity. Finally, a model experiment of detecting short oligonucleotides has demonstrated the biomedical application of this hyperspectral fluorescence biosensor.
