Nocturnal oximetry and transcutaneous carbon dioxide in home-ventilated neuromuscular patients
Julie Nardi1, Hélène Prigent, Annie Adala
1Services de Physiologie, Explorations Fonctionnelles, Réanimation Médicale et Centre d'Investigation Clinique et d'Innovation Technologique, Unité 805 Institut National de la Santé et de la Recherche Médicale, Hôpital Raymond Poincaré, Université de Versailles Saint-Quentin-en-Yvelines, Garches, France. jnardi@chu-reims.fr
Background:
Pulse oximetry alone has been suggested to determine which patients on home mechanical ventilation (MV) require further investigation of nocturnal gas exchange. In patients with neuromuscular diseases, alveolar hypoventilation (AH) is rarely accompanied with ventilation-perfusion ratio heterogeneity, and, therefore, oximetry may be less sensitive for detecting AH than in patients with lung disease.
Objective:
To determine whether pulse oximetry (S(pO(2))) and transcutaneous carbon dioxide (P(tcCO(2))) during the same night were interchangeable or complementary for assessing home MV efficiency in patients with neuromuscular diseases.
Methods:
Data were collected retrospectively from the charts of 58 patients with chronic neuromuscular respiratory failure receiving follow-up at a home MV unit. S(pO(2)) and P(tcCO(2)) were recorded during a 1-night hospital stay as part of standard patient care. We compared AH detection rates by P(tcCO(2)), S(pO(2)), and both.
Results:
AH was detected based on P(tcCO(2)) alone in 24 (41%) patients, and based on S(pO(2)) alone with 3 different cutoffs in 3 (5%), 8 (14%), and 13 (22%) patients, respectively. Using both P(tcCO(2)) and S(pO(2)) showed AH in 25 (43%) patients.
Conclusions:
Pulse oximetry alone is not sufficient to exclude AH when assessing home MV efficiency in patients with neuromuscular diseases. Both P(tcCO(2)) and S(pO(2)) should be recorded overnight as the first-line investigation in this population.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...