The effect of targeting scheme on tidal volume delivery during volume control mechanical ventilation
Teresa A Volsko1, Justin Hoffman, Alecia Conger
1Department of Health Professions, Youngstown State University, Youngstown, Ohio, USA. TVolsko@chmca.org
Background:
Technological advances have increased ventilator mode complexity and risk of operator error.
Objective:
To compare differences in volume control (VC) ventilation with set-point and dual targeting. Two hypotheses were tested: tidal volume (V(T)) delivery is different with VC using set-point versus dual targeting during active versus passive breathing; VC with dual targeting delivers V(T) similar to pressure support ventilation (PS) with active breathing.
Methods:
The Ingmar Medical ASL 5000 lung model simulated pulmonary mechanics of an adult patient with ARDS during active and passive ventilation. Resistance was standardized at 10 cm H(2)O/L/s and compliance at 32 mL/cm H(2)O. Active breathing was simulated by setting the frequency (f) = 26 breaths/min, and adjusting the muscle pressure (P(mus)) to produce a V(T) of 384 mL. VC was initiated with the Puritan Bennett 840 (set-point targeting) and the Servo-i (dual targeting) at V(T) = 430 mL, mandatory f = 15 breaths/min, and PEEP = 10 cm H(2)O. During PS, cycle threshold was set to 30% and peak inspiratory pressure adjusted to produce a V(T) similar to that delivered during VC. Expiratory V(T) was collected on 10 consecutive breaths during active and passive breathing with VC and PS. Mean V(T) differences (active vs passive model) were compared using analysis of variance. Statistical significance was established at P < .05.
Results:
The mean ± SD V(T) difference varied with targeting schemes: VC set-point = 37.3 ± 3.5 mL, VC-dual = 77.1 ± 3.3 mL, and PS = 406.1 ± 1.5 mL (P < .001). Auto-triggering occurred during VC set-point with the active model.
Conclusions:
Dual targeting during VC allows increased V(T), compared to set-point, but not as much as PS.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)