Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
Morphometric measurement of Schlemm's canal in normal human eye using anterior segment swept source optical coherence
Guohua Shi1, Fei Wang, Xiqi Li
1Chinese Academy of Sciences, The Key Laboratory on Adaptive Optics, Chengdu 610209, China. guohua_shi@yahoo.com.cn
Abstract:
We have used anterior segment swept source optical coherence tomography to measure Schlemm's canal (SC) morphometric values in the living human eye. Fifty healthy volunteers with 100 normal eyes were measured in the nasal and temporal side. Comparison with the published SC morphometric values of histologic sections proves the reliability of our results. The statistical results show that there are no significant differences between nasal and temporal SC with respect to their diameter, perimeter, and area in our study (diameter: t=0.122, p=0.903; perimeter: t=-0.003, p=0.998; area: t=-1.169, p=0.244); further, no significant differences in SC morphometric values are found between oculus sinister and oculus dexter (diameter: t=0.943, p=0.35; perimeter: t=1.346, p=0.18; area: t=1.501, p=0.135).