Related Experiment Video
Updated: May 24, 2026

Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Three-dimensional radiochromic film dosimetry of proton clinical beams using a gafchromic EBT2 film array
Jinsung Kim1, Myonggeun Yoon, Seonkyu Kim
1Department of Radiation Oncology, Samsung Medical Center, Seoul, Korea.
Abstract:
In this work, three-dimensional (3D) film-based proton beam measurements were used for the first time to verify the patient-specific radiation dose distribution, beam range and compensator shape. Three passively scattered proton beams and one uniform scanning proton beam were directed onto an acrylic phantom with inserted Gafchromic EBT films. The average gamma index for a comparison of the dose distributions was less than one for 97.2 % of all pixels from the passively scattered proton beams and 98.1 % of all pixels for the uniform scanning proton beams, with a 3 % dose and a 3 mm distance-to-dose agreement tolerance limit. The results also showed that the average percentage of points within the acceptance criteria for proton beam ranges was 94.6 % for the passively scattered proton beams. Both the dose distribution and the proton beam range determined by the 3D EBT film measurement agreed well with the planning system values.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022