Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Magnetization spoiling in radial FLASH contrast-enhanced MR digital subtraction angiography
Parmede Vakil1, Sameer A Ansari, Michael C Hurley
1Department of Biomedical Engineering, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.
Purpose:
To increase the in-plane spatial resolution and image update rates of 2D magnetic resonance (MR) digital subtraction angiography (DSA) pulse sequences to 0.57 × 0.57 mm and 6 frames/sec, respectively, for intracranial vascular disease applications by developing a radial FLASH protocol and to characterize a new artifact, not previously described in the literature, which arises in the presence of such pulse sequences.
Materials And Methods:
The pulse sequence was optimized and artifacts were characterized using simulation and phantom studies. With Institutional Review Board (IRB) approval, the pulse sequence was used to acquire time-resolved images from healthy human volunteers and patients with x-ray DSA-confirmed intracranial vascular disease.
Results:
Artifacts were shown to derive from inhomogeneous spoiling due to the nature of radial waveforms. Gradient spoiling strategies were proposed to eliminate the observed artifact by balancing gradient moments across TR intervals. The resulting radial 2D MR DSA sequence (2.6 sec temporal footprint, 6 frames/sec with sliding window factor 16, 0.57 × 0.57 mm in-plane) demonstrated small vessel detail and corroborated x-ray DSA findings in intracranial vascular imaging studies.
Conclusion:
Appropriate gradient spoiling in radial 2D MR DSA pulse sequences improves intracranial vascular depiction by eliminating circular banding artifacts. The proposed pulse sequence may provide a useful addition to clinically applied 2D MR DSA scans.